Select Page

HTM-ISO standards version

Technical Specifications

HTM-ISO TotalAlert Infinity Signal Capabilities
Master Version
(HTM/ISO)
Up to 20 onboard direct Line Contact Monitoring inputs
Area VersionsUp to 8 BeaconMedaes pressure transmitters
Combination VersionsCombinations of direct inputs, pressure transmitters and 4-20 mA inputs, limited by display configuration
Specialty MonitoringUp to 4 4-20 mA devices

Dimensions

HTM-ISO TotalAlert Embedded Network Specifications
BackboneStandard Ethernet
Number of DevicesLimited only by the network
Available DevicesMedipoint and other competitor networks and panels
Building Management InterfaceVia Ethernet BAC Net
CommunicationEach TotalAlert device can communicate via an SMTP gateway to send email and SMS texts to any receiving platform
Software RequiredAny standard web browser

Wireless capable via any method eligible for implementing wireless Ethernet TotalAlert Infinity® panels includes a wireless access point for setup and remote interrogation only (wireless control is not possible for any TotalAlert network device). Wireless can be disabled for security. Any alarm panel can be remotely observed on the network using a standard web browser.